Description
Standards
Conformity to standards | CE |
Certificate number and date | CEI/IC/G1/510/10.01.2019 |
Performance Parameters
Purpose | Separating Serum from blood or for urine sedimentation |
Centrifuge Design | Bench Top |
Weight including Default Rotor in kg | 18 |
Auto cut off | Yes |
Timer provided | Yes |
Maximum RPM | 2800 |
Number of steps for RPM variation | 7 |
Incremental Steps of Variable RPM | 400 |
Type of Motor | Brushless (without carbons) |
Maximum Noise level whemn working at 1 meter distance in db | 56 |
Ambient operating Temperature range in Degree centigrade Celsius and humidity range | 0 to 50 degree centigrade and Relative humidity 15 to 90% |
Rotor imbalance diagnostics and automatic stop if required | Yes |
Facility to ensure that In case Machine is runing and lid is opened with manual lock by user , machine shall stop | Yes |
Maximum Timer Set range | 60 minutes |
Timer settings Provided | interval of 10 secondsup to 1 minute and over 1 minute up to 60 minutes in interval of 1 minute |
Display | LCD |
Speed Controller | Microprocessor controlled pre programmed speed controller |
Speed Control Accuracy in case of set speed (Plus minus RPM ) | 50 |
Rotor configuration | Swing Out Rotor |
Rotor Feature with reference to Autoclavability | Autoclavable Rotor @ 121 Degree Celsus, |
Maximum Volume of the tube testable | 15 ml centrifuges tubes |
Rotor Size | 12x15ml, |
Power supply | Single phase, 230 V 50 HZ AC Source |
Power concumption in watts | 120 |
Length of the Medical Centrifuge in mm | 460 |
Width of the Medical Centrifuge in mm | 310 |
Height of of the Medical Centrifuge in mm | 178 |
Length of Power cord in meters | 3 |
The Equipment shall be installed and demonstrated before Acceptance installation and demonstration | No(No installation and demonstration required and consignee can install) |
Accessories | Spanner to tight or loose Nut of Rotor on Motor shaft head puller,spare rotor nuts ,fuse-2, Instruction Manual Book ,installation CD |
Warranty in Years | 3 |
Material Parameters
Assembly Material for manufacture of Medical Centrifuge | Consists of Die cast Aluminum, Aluminum sheet, Stainless Steel, CRC Steel,Plastic and Rubber |
Lid Material of Medical Centrifuge | C.R.C Steel Finish with Powder Coating or stove painting / ABS Plastic of injection moulding |
Body Material of Medical Centrifuge | C.R.C Steel Finish with Powder Coating or stove painting / ABS PLASTIC OF injection moulding |
Centrifuge Bowl material of Medical Centrifuge | 304 Quality Stainless Steel Moulded |
Material of Rotor | Aluminium, |
Reports
ISO certification for Manufacturer | Yes |
Certification number and date , if no certificate provided put NA | ISO-QMSI/1168/09.01.2019 |
Copies of all certifications and reports to be provided to buyer on demand at time of supplies | Yes |